CATALOG #: 90.P3900-180
The Pointe 180 impulse™ is an ISE Electrolyte Analyzer specifically designed for the direct determination of Na+, K+ and Cl- by ion selective electrodes in whole blood, serum, and aqueous QC material.
Reagent packs, electrodes, and the instrument itself all contain unique electronics allowing for self-calibration and monitoring for trouble shooting.
The analyzer is a bottle-free, stand alone, bench top, ISE electrolyte analyzer designed for the direct determination of Na+, K+ and Cl- by ion selective electrodes. The unit contains a built in thermal dot matrix printer. The unit is programmable through custom software allowing the user to define programs following their own established protocols. The software is displayed on an interactive touch screen which guides the user step-by-step through the process using on screen prompts, error messages and graphical displays.
Reagents are supplied in packs which plug into the side of the unit with no user contact to waste material. The unit will track reagent expiration and quantity of reagent remaining. *
* Not available for clinical human blood testing in the US
Catalog #: | Product: |
90.P3900-10 | Pointe 180 impulse™ |
Catalog #: | Product: |
90.B11-100 | Reagents |
90.B11-110 | Na+ Electrode |
90.B11-120 | K+ Electrode |
90.B11-130 | Cl- Electrode |
90.B11-140 | Reference Electrode |
90.B11-310 | Tri-Level QC Kit |
Method: | Direct measurement by ion selective electrode |
Sample Type: | Whole Blood, Serum, Aqueous QC Material |
Sample Size: | 60µl ± 5µl |
Sample Rate: |
48/hour maximum without printing 45/hour maximum with printing |
Measurement Ranges: |
Sodium — 40 - 205 mmol/L Potassium — 1.5 - 15.0 mmol/L Chloride — 50 - 200 mmol/L |
Sample Probe: | 316 Stainless Steel |
Vessels: | Capillary tubes with adaptor, tubes up to 100mm depth, syringes, sample cups |
Electrodes: |
Sodium (Na+) Electrode Potassium (K+) Electrode Chloride (Cl-) Electrode |
Analysis Time: | Typically < 60 seconds |
Patient Data Storage: | 200 results (with capability to assign a 15 character patient identifier) |
QC Data Storage: | 270 readings (30 readings each of 3 controls for each of the 3 electrodes, includes Levey-Jennings) |
Calibration: | On demand or automatic from 1 to 8 hour intervals |
Power Requirements: | 100 - 240VAC, 50 - 60Hz 90W maximum |
Dimensions: | 14”w x 12”h x 8”d (35.5cm x 30.5cm x 20cm) |
Weight (with reagent pack): | 14.5 lbs (6.6 kg) |
Display: | Interactive touch screen 3.5” (9cm) LCD color graphic display |
Interface: | Indirect via USB Drive (Included) |
Printer: | Thermal dot matrix with graphic capability, 29 character width |